BCI FingerPrint

Hand-Held Pulse Oximeter

Displays, Indicators, and Keys:
SpO2 LED numeric display
0.43 inch (10.9 mm) high
Pulse Rate LED numeric display
0.43 inch (10.9 mm) high
Pulse Strength Logarithmically scaled, 8 segment
bar graph
Keys On/Off, Print and Speaker On/Off


SpO2:
Range 0-99%
Accuracy +/- 2% at 70-99%; <=69% unspecified
Averaging 8 beat average


Pulse Rate:
Range 30-254 bpm
Accuracy +/- 2% at 30-254 bpm
Averaging 8 second average


Trend Storage:
Parameters SpO2, Pulse Rate and Relative Time
Interval 30 seconds
Capacity 14 hours of storage at 30-second intervals


Dimensions: 6.75" H x 3" W x 1.5" D
Weight: 13 ounces with batteries
Power Requirements: (4) "AA" Alkaline or NiCAD Batteries
Battery Life: 24 hours without printing
14 hours of continuous printing
spot checks
1 minute on
Intergrated Printer: 1-nch thermal printer paper
15 characters per line
Printer Output: RS232C


Mechanical Durability Test Results:
Vibration Testing In compliance with IEC 68-2-34, IEC 68-2-6,
and MIL-STD-810E, Method 514.4
Shock Testing In compiance with IEC 68-2-27
Drop Testing IEC 601-1, section 21.5
Environmental Specifiacations:
Operating Temperature 32oF to 131oF(0oC to 55oC)
Storage Temperature -40oF to 167oF(-40oC to 75oC)
Relative Humidity
(non-condensing)
10-95% (storage)
15-95% (operating)

To join the network today